ICell box

ICell box (Enlarge)

1. The 12 chambers and sample tubes should be dry to avoid tube sticking upon freezing.

2. Make sure the core (metal ring) is at room temperature and seated in the bottom of the central cavity.

3. Place sample vials containing maxinum 2 ml of cell suspension in each well .The sample vials should not extend about the iCellbox body.

4. Check that the tubes slide in and out freely.

5. Fully seat the lid on the iCellbox.

6. Put the iCellbox upright into a -80℃ freezer or dry ice locker. Ensure that there is at least 1 inch of free space clearance around the iCellbox.

7. Freeze for 4 hours before transferring samples to archive storage.